Proceedings of the Human Factors and Ergonomics Society Annual Meeting · 2004 · 19 citations · 5 references
EngineeringMechanical EngineeringMechatronics EngineeringErgonomicsHuman PerformanceMechanical DesignInstrument HandlingLaparoscopic InstrumentsActuation ForceInstrument Handle DesignLaparoscopic Instrument Handling
Twenty-two subjects were tested and categorized according to hand size (small, medium, or large). Each subject selected the best location for a trackball and a trigger on a handle. Each subject specified the optimum diameter/size of the handle that he or she preferred. Additionally, subjects selected their preferred pivot range for opening and closing the handle. Finally, each subject exerted his or her preferred force for the trackball and trigger controls in the selected positions. Based on the data collected in this experiment, the recommended handle diameter is in the range of 4.3 to 5.7 cm. The recommended handle pivot is the range of 8.1 to 17.3 degrees for the open and closed positions. The recommended trackball actuation force is 3.0 lbs and the recommended ratchet actuation force is 0.6 lbs, on average.
5
Ergonomic problems associated with laparoscopic surgery
Ramón Berguer, D. L. Forkey, Warren D. Smith · Surgical Endoscopy · 1999 · 357 citations
Engineering, Surgery, Ergonomics +12
Surgical technology and the ergonomics of laparoscopic instruments
Ramón Berguer · Surgical Endoscopy · 1998 · 208 citations
Nursing, Surgery, Ergonomics +13
Hand Anthropometry of U.S. Army Personnel
Thomas M. Greiner · 1991 · 126 citations