Calibration Curve for Quantitative Analysis of the Monoclinic-Tetragonal ZrO<sub>2</sub>System by X-Ray Diffraction
815
Citations
14
References
1984
Year
Materials ScienceX-ray CrystallographyEngineeringNonlinear Calibration CurveCalibrationX-ray DiffractionApplied PhysicsQuantitative AnalysisX-ray Powder Pattern-fittingVolume Fraction VsCrystallographyCalibration Curve
A nonlinear calibration curve for volume fraction vs integrated intensity ratio was obtained for the monoclinic-tetragonal ZrO2 system by using X-ray powder pattern-fitting and pattern-decomposition techniques. The empirical equation agrees well with the theoretical one. By using this equation, the deviation from linearity (6.8% maximum) resulting from the intensity difference of corresponding reflections of the two phases can be estimated quite accurately.