Basic Performance Test of a Prototype PET Scanner Using CdTe Semiconductor Detectors

Yuuichirou Ueno, Yuichi Morimoto, Katsutoshi Tsuchiya, Norihito Yanagita, Soichi Kojima, Takafumi Ishitsu, Hiroshi Kitaguchi, Naoki Kubo, Shen Zhao, Nagara Tamaki,

IEEE Transactions on Nuclear Science · 2009 · 18 citations · 8 references

Concepts

Abstract

A prototype positron emission tomography (PET) scanner using CdTe semiconductor detectors was developed, and its initial evaluation was conducted. The scanner was configured to form a single detector ring with six separated detector units, each having 96 detectors arranged in three detector layers. The field of view (FOV) size was 82 mm in diameter. Basic physical performance indicators of the scanner were measured through phantom studies and confirmed by rat imaging. The system-averaged energy resolution and timing resolution were 5.4% and 6.0 ns (each in FWHM) respectively. Spatial resolution measured at FOV center was 2.6 mm FWHM. Scatter fraction was measured and calculated in a National Electrical Manufacturers Association (NEMA)-fashioned manner using a 3-mm diameter hot capillary in a water-filled 80-mm diameter acrylic cylinder. The calculated result was 3.6%. Effect of depth of interaction (DOI) measurement was demonstrated by comparing hot-rod phantom images reconstructed with and without DOI information. Finally, images of a rat myocardium and an implanted tumor were visually assessed, and the imaging performance was confirmed.