A Robust Uniaxial Force Sensor for Minimally Invasive Surgery
95
Citations
10
References
2010
Year
EngineeringDevice TherapyMinimally Invasive ProcedureMechanical EngineeringSurgeryBiomedical EngineeringSoft RoboticsBiomechanicsMinimally Invasive SurgeryCardiologyCardiac MechanicCardiovascular ImagingMechanobiologyImplantable SensorMechanical Circulatory SupportMitral Valve AnnuloplastyFiber Optic TransductionBeating HeartMedicine
This paper presents a novel miniature uniaxial force sensor for use within a beating heart during mitral valve annuloplasty. The sensor measures 5.5 mm in diameter and 12 mm in length and provides a hollow core to pass instrumentation. A soft elastomer flexure design maintains a waterproof seal. Fiber optic transduction eliminates electrical circuitry within the heart, and acetal components minimize ultrasound-imaging artifacts. Calibration uses a nonlinear viscoelastic method, and in vitro tests demonstrate a 0-4-N force range with rms errors of 0.13 N (< 3.2%). In vivo tests provide the first endocardial measurements of tissue-minimally invasive surgery instrument interaction forces in a beating heart.