On the Calibration of Direct-Current Current Transformers (DCCT)

Luca Callegaro, Cristina Cassiago, E. Gasparotto

IEEE Transactions on Instrumentation and Measurement · 2014 · 30 citations · 5 references

DOIFull text

Open access

Concepts

Abstract

Modern commercial dc current transformers (DCCTs) can measure currents up to the kiloampere range with accuracies better than 1 × 10 <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">-5</sup> . We discuss here a DCCT calibration method and its implementation with commercial instruments typically employed in low-resistance calibration laboratories. The primary current ranges up to 2 kA; in the current range <;100 A, the calibration uncertainty is better than 3 × 10 <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">-7</sup> . An example of calibration of a high-performance DCCT specified for primary currents measurement up to 900 A is discussed in detail.