IEEE Transactions on Instrumentation and Measurement · 2014 · 30 citations · 5 references
EngineeringElectrical EngineeringAnalog ElectronicsInstrumentationMetrologyCalibrationCurrent Measurement InstrumentsElectrical MeasurementElectrical MetrologyEquipment CalibrationSensor CalibrationCalibration UncertaintyCalibration Uncertainty AnalysisCurrent MeasurementCurrent Measurement AccuracyCurrent Measurement UncertaintyCurrent SensingCurrentsUncertainty Calibration
Modern commercial dc current transformers (DCCTs) can measure currents up to the kiloampere range with accuracies better than 1 × 10 <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">-5</sup> . We discuss here a DCCT calibration method and its implementation with commercial instruments typically employed in low-resistance calibration laboratories. The primary current ranges up to 2 kA; in the current range <;100 A, the calibration uncertainty is better than 3 × 10 <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">-7</sup> . An example of calibration of a high-performance DCCT specified for primary currents measurement up to 900 A is discussed in detail.
5
A novel current calibration system up to 20 ka
G. Fernqvist, Björn Halvarsson, J. Pett et al. · IEEE Transactions on Instrumentation and Measurement · 2003 · 21 citations · Full text
P. Odier · Prepared for · 2004 · 12 citations