2008 · 10 citations · 4 references
EngineeringBiomedical EngineeringElectrical EngineeringElectrophysiologyNeural EngineeringNeurophysiologyImplantable Biomedical DevicesImplantable Neural DevicesNeural InterfacesNeural Signal AcquisitionNeural StimulationNeurostimulationBicmos TechnologyImplantable Neural ElectrodesImplantable Neural StimulationNeural ElectrodesNeural Stimulation DevicesNeural Stimulation Through Electrode ArraysSciatic Nerve StimulationWireless Neural Interfaces
We present the design of an integrated circuit for wireless neural stimulation, along with bench-top and in-vivo experimental results. The chip has the ability to drive 100 individual stimulation electrodes with constant-current pulses of varying amplitude, duration, interphasic delay, and repetition rate. The stimulation is done using a biphasic (cathodic and anodic) current source, injecting and retracting charge from the nervous system. Wireless communication and power are achieved over a 2.765-MHz inductive link. Only two off-chip components are needed to operate the stimulator: a 10-nF capacitor to aid in power supply regulation and a coil for power and command reception. The chip was fabricated in a commercially available 0.6-mum 2P3M BiCMOS process. The chip was able to activate motor fibers to produce muscle twitches via a Utah Slanted Electrode Array implanted in cat sciatic nerve, and to activate sensory fibers to recruit evoked potentials in somatosensory cortex.
4