Robust beat detector for ambulatory cardiac monitoring
50
Citations
10
References
2009
Year
Unknown Venue
Robust Beat DetectionElectrophysiological EvaluationPhysical ActivityEngineeringMeasurementBiosignal ProcessingWearable TechnologyComputer EngineeringPatient MonitoringRobust Beat DetectorCardiologyBiostatisticsHealth MonitoringElectrophysiologySignal ProcessingWearable SensorBeat Detection AlgorithmHealth Sciences
Robust beat detection under noisy conditions is required in order to obtain a correct clinical interpretation of the ECG in ambulatory settings. This paper describes the evaluation and optimization of a beat detection algorithm that is robust against high levels of noise. An evaluation protocol is defined in order to study four different characteristics of the algorithm: non-rhythmic patterns, different levels of SNR, exact peak detection and different levels of physical activity. This protocol is based on the MIT/BIH arrhythmia database and additional ECG recordings obtained under different levels of physical activity measured by 2-axis accelerometers. The optimized algorithm obtained a Se=99.65% and +P=99.79% on the MIT/BIH arrhythmia database while keeping a good performance on ECGs with high levels of activity (overall of Se=99.86%, +P=99.91%). In addition, this method was optimized to work in real time, for future implementation in a Wireless ECG sensor based on a microprocessor.