Astronomy and Astrophysics · 2010 · 42 citations · 21 references
EngineeringAstronomyInstrumentationMetrologySpectroscopyCalibrationEquipment CalibrationInstrument ValidationNoise MeasurementSpectral CalibrationFlight OperationsFrequency CalibrationInstrument CalibrationInstrument Noise CharacterizationInstrument Performance CharacterizationInstrumentation CalibrationPre-launch CalibrationScientific Performance Metrics
We present the calibration and scientific performance parameters of the <i>Planck<i/> Low Frequency Instrument (LFI) measured during the ground cryogenic test campaign. These parameters characterise the instrument response and constitute our optimal pre-launch knowledge of the LFI scientific performance. The LFI shows excellent 1/<i>f<i/> stability and rejection of instrumental systematic effects; its measured noise performance shows that LFI is the most sensitive instrument of its kind. The calibration parameters will be updated during flight operations until the end of the mission.
21
<i>Planck</i>pre-launch status: The<i>Planck</i>-LFI programme
N. Mandolesi, M. Bersanelli, R. C. Butler et al. · Astronomy and Astrophysics · 2009 · 122 citations · Full text
Natural Sciences, Astronomy, Physics +16
<i>Planck</i>pre-launch status: The optical system
J. A. Tauber, H. U. Nørgaard-Nielsen, P. A. R. Ade et al. · Astronomy and Astrophysics · 2010 · 86 citations · Full text
Natural Sciences, Astronomy, Physics +17