Miniature, high-resolution, quadrupole mass-spectrometer array

O. J. Orient, A. Chutjian, Vachik Garkanian

Review of Scientific Instruments · 1997 · 90 citations · 2 references

Concepts

TL;DR

The array uses 16 rods (25 mm × 2 mm) with a miniature Nier-type ionizer, a channel-type electron multiplier detector, and operates at 5.3, 7.1, and 12.9 MHz. The 4×4 array achieves a mass range of 1–300 u, a resolution of 0.1–0.5 u (m/Δm ≈ 600), and a sensitivity of ~1×10¹² counts/Torr s.

Abstract

A miniature quadrupole mass spectrometer array consisting of 16 rods in a 4×4 array is reported. Each rod is 25 mm in length and 2 mm in diameter. The ionizer is of a miniature Nier-type design, and the detector is a channel-type electron multiplier. Operating frequencies are 5.3, 7.1, and 12.9 MHz. The mass range demonstrated herein is 1–300 u; and the resolution of the system is 0.1–0.5 u (full width at half-maximum), or m/Δm=600. The present sensitivity is calculated and measured to be approximately 1×1012 counts/Torr s.