Spectrally-sampled OCT for sensitivity improvement from limited optical power

Eun Joo Jung, Jaeseok Park, Myung Yung Jeong, Chang‐Seok Kim, Tae Joong Eom, Bong-Ahn Yu, S. Gee, Jong-Min Lee, Moon Ki Kim

Optics Express · 2008 · 22 citations · 17 references

DOIFull text

Open access

Concepts

Abstract

Although high optical illumination power is favored in optical coherence tomography (OCT) for better signal-to-noise ratio, optical power is often limited by a damaged threshold for biomedical living tissues and autocorrelation signals observed in tomograms. In order to improve signal sensitivity without increasing the optical illumination power, a spectrally sampled multi-wavelength light source is proposed for the OCT system. A fiber Sagnac comb filter was used to spectrally sample the output of a continuous spectral light source. Point spread function analysis shows that the spectrally sampled OCT has an almost 50% dynamic range improvement in comparison with a conventional continuous spectral light source OCT for the same average optical power of 6 mW.

References

17