Ionization chamber gradient effects in nonstandard beam configurations

85

Citations

27

References

2009

Year

Abstract

To reduce or eliminate the magnitude of perturbation correction factors in nonstandard beam reference dosimetry, two possible ways to report absorbed dose are suggested: (1) Reporting average dose to the sensitive volume of the chamber filled with water, combined with removing the reference field implicit gradient effect when measuring output factors, and (2) reporting average dose to the chamber itself during output factor verifications. The first option could be adopted if clinical beam correction factors are negligible. The second option could simplify quality assurance procedures when correction factors are not negligible and have to be calculated using Monte Carlo simulations.