Electrophoresis · 2001 · 38 citations · 23 references
PharmacyPharmaceutical AnalysisPharmacologyAnalytical ChemistryAnalytical ToxicologyMass Spectrometric AnalysisBioanalytical Method ValidationChromatographyContaminant AnalysisInstrumental AnalysisLiquid ChromatographyMetabolite ExtractionAssay DeterminationBioanalytical ChromatographyCalibration CurvesCapillary ElectrophoresisCapillary Electrophoresis Method DevelopmentConcentrationDetection Limit AssessmentMethod Detection Limit
In this study, low concentrations of histamine2-receptor (H2-)antagonists were effected across a water plug, with separation taking place in a binary buffer comprising ethylene glycol and NaH2PO4 (pH 5.0), and detection at 214 nm. Liquid-liquid extraction with ethyl acetate- isopropanol is shown to provide extracts that are sufficiently clean. The calibration curves were linear over a concentration range of 0.1-2.00 microg/mL cimetidine, 0.2-5.0 microg/mL ranitidine-HCl, 0.3-5.0 microg/mL nizatidine, and 0.1-3.0 microg/mL famotidine. Mean recoveries were > 82%, while the intra- and interday relative standard deviations (RSDs) and relative errors (REs) were all < 13%. The method is sensitive with a detection limit of 3 ng/mL cimetidine, 30 ng/mL ranitidine HCl, 50 ng/mL nizatidine and 10 ng/mL famotidine (S/N = 3, electric-driven injection 90 s). This newly developed capillary electrophoresis (CE) method was applied for the determination of analytes extracted from plasma taken from a volunteer dosing a cimetidine, ranitidine, and nizatidine tablet simultaneously. These three H2-antagonists can be detected in real samples by this method, excluding the low dosing of famotidine tablet.
23