Publication | Closed Access
Implant Surface Analysis and Microbiologic Evaluation of Failed Implants Retrieved From Smokers
30
Citations
39
References
2007
Year
Failed Implant SurfacesElectron MicroscopyMicrobiologic EvaluationFailed ImplantsImplantologyImplant RemovalDental BiomechanicsOral MicrobiologyMicrobiologyImplantable DeviceMedicineImplant Surface Analysis
The aim of this study was to evaluate the microbiota and surface of failed titanium dental implants from 4 manufacturers. Twelve mobile dental implants were retrieved from 10 smokers after 3 to 10 years of functional loading. Before implant removal, microbial samples were taken and evaluated using polymerase chain reaction. After implant removal, analyses of the failed implant surfaces were performed using scanning electron microscopy and energy-dispersive spectrometer x-ray. Periodontal pathogens such as Aggregactibacter actinomycetemcomitans, Campylobacter rectus, Eikenella corrodens, Fusobacterium nucleatum, Porphyromonas gingivalis, Prevotella intermedia, Tannerella forsythia, and Treponema denticola were detected in all implants in different proportions. Surface analysis showed varying degrees of surface roughness between the samples and the presence of proteinaceous material, appearing mainly as dark stains. Foreign carbon, oxygen, sodium, calcium, aluminum, and silicon elements were also found. Although no material-related causes of implant failure were detected, several periodontal pathogens were identified independently of the surface topography or manufacturer.