Channeltron efficiency for counting of H+ and H− at low energy

D. H. Crandall, J. A. Ray, Carmen Cisneros

Review of Scientific Instruments · 1975 · 84 citations · 12 references

Concepts

Abstract

The counting efficiency of a channel electron multipier is measured for H+ and H− impact with the input of the Channeltron grounded and for impact energies between 0.3 and 5 keV. The counting efficiency is found to be about 30% greater for H− than for H+ at energies below 1 keV. This difference and an observed difference between H− and H+ in the variation of counting efficiency across the Channeltron opening are attributed to the surface potential of the Channeltron near the opening, which can accelerate or decelerate the incident particles.

References

12