IEEE Transactions on Instrumentation and Measurement · 1989 · 13 citations · 2 references
EngineeringElectrical EngineeringInstrumentationMicrowave EngineeringTest ConstructionCalibrationDevice CharacterizationElectrical MeasurementImpedance MatchingImpedance Matching CircuitsMicrowave MeasurementsCalibration ProceduresCalibration StandardsImpedance MeasurementImpedance ParametersRadio Frequency Analysis
The test ports on automatic network analyzers are generally built with an impedance that matches the impedance of the calibration standards. Experimental evidence is presented that substantial impedance discontinuities can be tolerated at the test port interface if proper calibration procedures are observed. The 50- Omega test port on one of the six-ports in a dual six-port network analyzer was replaced with a 75- Omega test port. This test port was then calibrated to look like a 50- Omega test port. Measurements on various devices show that it is possible to make a 75- Omega test port indistinguishable from a 50- Omega test port.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">></ETX>
2