MedicineCardiologyRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyCardiac ImagingCardiothoracic ImagingCardiovascular ImagingComputed TomographyImage ReconstructionMedical ImagingCardiac Computed TomographyData ReconstructionGlobal Ventricular FunctionHeart RateQuantification Of Cardiac ChambersQuantitative Cardiac ImagingTomographic AcquisitionVentricular Ejection Fraction Assessment
PURPOSE: To investigate the influence of heart rate and temporal resolution on the assessment of global ventricular function with dual-source computed tomography (DSCT). MATERIALS AND METHODS: A dynamic cardiac phantom was repeatedly scanned with a DSCT scanner applying a standardized scan protocol at different heart rates, ranging from 40 to 140 bpm. Images were reconstructed with monosegmental and bisegmental algorithms using data from a single source and from both sources. Ventricular volumes and ejection fraction (EF) were computed by semiautomated analysis. Results were compared with the phantom's real volumes. Interscan, intraobserver, and interobserver variability were calculated. RESULTS: For single-source data reconstruction temporal resolution was fixed to 165 milliseconds, whereas dual-source image reconstructions resulted in a temporal resolution of 83 milliseconds (monosegmental) and 67.7+/-14.2 milliseconds (bisegmental), respectively. In general, deviation from the phantom's real volumes was less with dual-source data reconstruction when compared with single-source data reconstruction. Comparing dual-source data reconstruction with single-source data reconstruction, the percent deviation from the phantom's real volumes for EF was 0.7% (monosegmental), 0.7% (bisegmental), and 4.3% (single source), respectively. There was no correlation between heart rate and EF for dual-source data reconstruction (r=-0.168; r=-0.157), whereas a relevant correlation was observed for single-source data reconstruction (r=-0.844). Interscan, intraobserver, and interobserver variability for EF were 1.4%, 0.9%, and 0.3%, respectively. CONCLUSIONS: DSCT allows reliable quantification of global ventricular function independent of the heart rate. Multisegmental image reconstruction is not needed for DSCT assessment of global ventricular function.
29
First performance evaluation of a dual-source CT (DSCT) system
Thomas Flohr, Cynthia H. McCollough, Herbert Bruder et al. · European Radiology · 2005 · 1.4K citations
Mary J. Emond, Michael B. Mock, K B Davis et al. · Circulation · 1994 · 556 citations · Full text
Medicine, Cardiology, Surgery +10
Radiation Exposure during Cardiac CT: Effective Doses at Multi–Detector Row CT and Electron-Beam CT
Peter Hunold, Florian Vogt, Axel Schmermund et al. · Radiology · 2003 · 408 citations
Medicine, Cardiology, Radiology +17
Dual-source CT cardiac imaging: initial experience
Thorsten R. C. Johnson, Konstantin Nikolaou, Bernd J. Wintersperger et al. · European Radiology · 2006 · 343 citations
Medicine, Cardiology, Radiology +13