Journal of Applied Physics · 1959 · 33 citations · 5 references
Natural SciencesMaterials EngineeringMaterials ScienceCrystal StructureMaterials CharacterizationRadiation MeasurementCrystallographyX-ray CrystallographyX-ray Data AnalysisX-ray DiffractionX-ray OpticsCrystal Size AnalysisCrystallite Size AnalysisCrystallite Size DeterminationExperimental Data CollectionExperimental Study DesignPowder DiffractionPowder X-ray DiffractionSilicon
The relative rms deviation σ of the intensity of a rapidly rotating polycrystallite specimen is σ=[6.5R sinθ]/[h(mNeff)½], where R=goniometer radius, θ=Bragg angle, h=(hf+hs)/2, hf=length of focal spot, hs=length of receiving slit, m=multiplicity factor, and Neff the effective number of irradiated crystallites. Experimental data on silicon powder specimens are presented to show the dependence of σ on crystallite sizes, choice of x-ray wavelengths and various other experimental parameters. Comparisons of data for rapidly rotating and stationary specimens having crystallite sizes in the 30–50 μ range show that rotation improves σ by a factor of 7 or 8 for peak intensities and 4 or 5 for integrated intensities, using a standard diffractometer.
5
Use of Counter Tubes in X-Ray Analysis
W. Parrish, Th. Köhler · Review of Scientific Instruments · 1956 · 86 citations
Statistical factors in X-ray intensity measurements
M. Mack, N. Spielberg · Spectrochimica Acta · 1958 · 32 citations