Fixed-Point Technique for Approximating the ITS-90 between 420°C and 1085°C with an Infrared Thermometer

M. Battuello, F Lanza, T Ricolfi

Metrologia · 1990 · 20 citations · 9 references

Concepts

Abstract

A fixed-point technique for calibrating infrared thermometers between the Zn (420°C) and Cu (1085°C) points was analyzed. To do this, a precision silicon detector thermometer working at 900 nm and 950 nm (band-widths of 70 nm and 40 nm, respectively) and a set of four fixed-point black-body cavities (Zn, Al, Ag and Cu) were used. The best results showed that the ITS-90 can be approximated to within ±0.05°C throughout the Zn-Cu interval when the thermometer in its 900 nm configuration is calibrated at the Zn, Al and Cu points.

References

9