Physics in Medicine and Biology · 2004 · 157 citations · 14 references
Computed TomographyImage ReconstructionEngineeringBiomedical EngineeringTreatment VerificationImage AnalysisCt ScanComputational ImagingComputational GeometryRadiologyHealth SciencesGeometric ModelingReconstruction TechniqueMedical ImagingFbp AlgorithmMedical Image ComputingMultislice Spiral CtNew 3DBiomedical ImagingSmpr AlgorithmImage QualityArbitrary Pitch3D Imaging
A new 3D reconstruction scheme, weighted filtered backprojection (WFBP) for multirow spiral CT based on an extension of the two-dimensional SMPR algorithm is described and results are presented. In contrast to other 3D algorithms available, the algorithm makes use of all available data for all pitch values. The algorithm is a FBP algorithm: linear convolution of the parallel data along the row direction followed by a 3D backprojection. Data usage for arbitrary pitch values is maintained through a weighting scheme which takes into account redundant data. If proper row weighting is applied, the image quality is superior to the image quality of the SMPR algorithm.
14
L.A. Feldkamp, L. C. Davis, J. W. Kress · Journal of the Optical Society of America A · 1984 · 6.2K citations
A general cone-beam reconstruction algorithm
Ge Wang, T. H. Lin, Ping Cheng et al. · IEEE Transactions on Medical Imaging · 1993 · 331 citations
Image reconstruction and image quality evaluation for a dual source CT scanner
Thomas Flohr, Herbert Bruder, Karl Stierstorfer et al. · Medical Physics · 2008 · 166 citations · Full text
Advanced single‐slice rebinning in cone‐beam spiral CT
Marc Kachelrieß, S. Schaller, Willi A. Kalender · Medical Physics · 2000 · 164 citations