X-ray diagnostic calibration with the tabletop laser facility EQUINOX

Charles Reverdin, Mathieu Paurisse, T. Caillaud, P. Combis, Alain Duval, D. Gontier, D. Husson, C. Rubbelynck, C. Zuber

Review of Scientific Instruments · 2008 · 16 citations · 15 references

Concepts

Abstract

The broadband x-ray emission of a target irradiated by a laser can be used to check the calibration of detectors. At CEA-DIF we have a tabletop picosecond laser facility called EQUINOX with 0.3 J at 800 nm. The laser is focused inside a target chamber onto a solid target and produces bright radiation in the 100-2000 eV spectral range. The x-ray source is routinely monitored with a pinhole camera for source dimension measurement and with x-ray diodes for flux measurement. In addition an x-ray transmission grating spectrometer, a crystal spectrometer, and a single count charge coupled device camera measure the x-ray spectrum between 100 eV and 15 keV. The absolute calibration of those sets of spectrometers allows us to fully characterize x-ray emission spectra. Typical duration is less than 100 ps. The spectrum can be tuned by changing target material, pulse length, and x-ray filters. An application to checking the calibration of x-ray diodes used in the broad band spectrometer DMX with single shots will be presented.