Journal of Physics E Scientific Instruments · 1988 · 18 citations · 3 references
EngineeringElectrical EngineeringAnalog ElectronicsMicroelectronicsSensorsAnalog InstrumentationCapacitive SensingElectrical CharacterizationElectrical MeasurementTransducer PrinciplesTransducer TechnologyCapacitance MeasurementCapacitance Measurement TechniquesCapacitance ModelingCapacitive SensorsFrequency MeasurementMulticomponent Flow ModelingTransducer DesignVoltage
An absolute capacitance measuring device based on the fundamental charge, capacitance and voltage relationship is described. It consists of a pair of CMOS analogue switches and a current integration amplifier, resulting in a simple capacitance transducer capable of providing a programmable measurement frequency ranging from 100 kHz to 5 MHz. This makes the transducer suitable for component concentration measurement of multicomponent flow processes involving conductive fluids. The initial circuit design is followed by a mathematical model of the transducer. Parametric analysis of this model is used to optimise both the sensitivity and stability of the transducer. A discrimination of 0.01 pF in 10 pF full scale is achieved. In measurements of capacitance with one plate grounded, the problem of stray capacitance arises. A novel switched guard method has been devised which reduces the effect of strays by a factor of 0.0025 at an operating frequency of 1 MHz.
3
Alternating-Current Bridge Methods
Students Quarterly Journal · 1940 · 197 citations