Journal of Applied Crystallography · 1981 · 16 citations · 0 references
Natural SciencesChemistryAnalytical ChemistryInstrumentationRadiation MeasurementCrystallographyX-ray CrystallographyX-ray Data AnalysisX-ray DiffractionDiffraction Data CollectionIntensity FluctuationsIonization Chamber DesignIonization Chamber PhysicsProtein CrystallographySynchrotron CrystallographyTime-resolved Crystallography
Time-variable incident intensity from synchrotron radiation sources may cause scaling errors in crystallographic data collection using integrative techniques as in rotation methods. This problem is avoided with a new technique which has been developed for computer-controlled film data collection with an Arndt–Wonacott camera. Effects of intensity fluctuations are automatically compensated by hardware on the basis of incident-beam monitoring with a small ionization chamber placed between collimator and crystal.