Review of Scientific Instruments · 2010 · 12 citations · 6 references
EngineeringNuclear EngineeringPhysicsInstrumentationNuclear InstrumentationPlasma PhysicsCalibrationEquipment CalibrationHigh-temperature PlasmasPlasma DiagnosticsSensor CalibrationTemperature MeasurementNoise TemperaturePower OutputPower Output ModelingRadiometer Calibration
We developed and installed an electron cyclotron emission radiometer for taking measurements of Korea Superconducting Tokamak Advanced Research (KSTAR) plasma. In order to precisely measure the absolute value of electron temperatures, a calibration measurement of the whole radiometer system was performed, which confirmed that the radiometer has an acceptably linear output signal for changes in input temperature. It was also found that the output power level predicted by a theoretical calculation agrees with that obtained by the calibration measurement. We also showed that the system displays acceptable noise-temperature performance around 0.23 eV.
6