British Journal of Radiology · 1952 · 29 citations · 7 references
MedicineDermatologyRadiation DosimetryRadiation ProtectionRadiobiologyBasal Cell CarcinomaRadiation TherapyRadiobiological ConsiderationsRadiotherapy For Skin CancerBeta Radiation DosimetryDensityDepth Dose DistributionRadiation Applicator SelectionSurface Dose Measurement
Radioactive phosphorus incorporated into thin polythene plastic for use as a β-ray applicator in superficial therapy is described. Measurements made with a shallow ionization chamber show surface dosage rates of the order of 1500 to 2000 r/hour. Depth dose measurements were made in unit density material and compared with similar measurements made with a 45 kVp X-ray plant. The protection problems are discussed and safe methods of handling and preparation of the applicators are described. The initial response to this β therapy of five patients with superficial rodent ulcers, and two patients with psoriasis, is presented. Preliminary clinical investigations show that the lesions to be treated must be carefully selected. The method appears applicable in particular to patients with multiple basal cell carcinomata.
7
The Clinical Use of Radioactive Isotopes.
&NA; · The American Journal of the Medical Sciences · 1950 · 33 citations
Dose Measurement in Beta-ray Therapy
G. W. Blomfield, F. W. Spiers · British Journal of Radiology · 1946 · 17 citations