Journal of Liquid Chromatography · 1994 · 26 citations · 18 references
PharmacyChemistryAnalytical ChemistryChromatographic AnalysisSeparation ScienceAnalytical ChromatographyChromatographic SeparationChromatographyMobile Phase OptimizationPka DeterminationStationary Phase ChemistryChromatographic Separation MechanismsMobile Phase Composition EffectsMobile Phase EffectsPh Effects On SeparationPka ValuesReversed-phase ChromatographyReversed-phase Liquid ChromatographyReversed-phase Mobile Phase Optimization
Abstract The dependence of retention on pH of the mobile phase is investigated in reversed-phase ion interaction chromatography, being the mobile phase an aqueous solution of 5.0 mM octylammonium and ortho-phosphoric acid at different pH values. The analytes considered were amines, diamines, amides, and species characterized by different functionalities. The behaviour of retention as a function of pH is discussed with comparison with literature data and correlation with pka values is shown. The optimization of separation and resolution in a mixture of acids and amines through pH variations is also presented.
18