Enhancement of Sensitivity for Antimony Determination in Atomic Absorption Spectrophotometry by Introducing Stibine into an Argon-Hydrogen Flame

Yoshinari Yamamoto, Takahiro Kumamaru, Yasunori Hayashi, Ryuzo Tsujino

Analytical Letters · 1972 · 18 citations · 1 references

Concepts

Abstract

Abstract The sensitivity of antimony determination with atomic absorption spectrophotometry is enhanced to a large extent by introducing stibine gas into an argon-hydrogen flame. As a reducing agent, zinc tablet made from the zinc powder paste with water is successfully used for quantitative and rapid productions of stibine from antimony (III) solution. The sensitivity for 1% absorption of the signal is estimated to be about 0.004 ppm of antimony. Key Words: antimonystibineatomic absorptionzinc tabletargon-hydrogen flame