Concepedia

Publication | Open Access

Reproducibility of viscoelastic property and intraocular pressure measurements obtained with the Ocular Response Analyzer

29

Citations

21

References

2010

Year

Abstract

The ORA device provides reproducible information on viscoelastic properties of the cornea in normal eyes notably CRF and CH. IOPcc was less reproducible. Four measurements per eye were necessary to reach a 10% precision and six for 5%.

References

YearCitations