Publication | Open Access
Densely folded spectral images of the CCD spectrometer working in the full 200-1000nm wavelength range with high resolution
29
Citations
0
References
2005
Year
High ResolutionEngineeringChemistryCcd SpectrometerSpectrochemical AnalysisSpectroscopic PropertyDispersion DirectionNew Charge-coupled DeviceOptical PropertiesSystem CalibrationInstrumentationFolded Spectral ImagesPhotonicsPhysicsImaging SpectroscopySpectral ImagingNatural SciencesSpectroscopyApplied PhysicsSpectral SearchingOptoelectronicsSpectroscopic Method
A new charge-coupled device (CCD) spectrometer has been studied and constructed by using a two-dimensional CCD detector and an integrated grating consisting of 10 subgratings. Effective spectral images of 268 mm along the dispersion direction have been densely folded 10 times to cover the full 200-1000 nm working wavelength range without any mechanical moving elements. The results show that the system has a spectral resolution and acquisition time of better than 0.07 nm and less than 100 ms, respectively, in the entire spectral range after system calibration.