A High-Throughput Time-Resolved Mini-Silicon Photomultiplier With Embedded Fluorescence Lifetime Estimation in 0.13 <formula formulatype="inline"> <tex Notation="TeX">$\mu$</tex></formula>m CMOS

David Tyndall, B. Rae, David Li, Jochen Arlt, A.H. Johnston, Justin Richardson, Robert Henderson

IEEE Transactions on Biomedical Circuits and Systems · 2012 · 83 citations · 23 references

Concepts

Abstract

We describe a miniaturized, high-throughput, time-resolved fluorescence lifetime sensor implemented in a 0.13 m CMOS process, combining single photon detection, multiple channel timing and embedded pre-processing of fluorescence lifetime estimations on a single device. Detection is achieved using an array of single photon avalanche diodes (SPADs) arranged in a digital silicon photomultiplier (SiPM) architecture with 400 ps output pulses and a 10% fill-factor. An array of time-to-digital converters (TDCs) with ≈50 ps resolution records up to 8 photon events during each excitation period. Data from the TDC array is then processed using a centre-of-mass method (CMM) pre-calculation to produce fluorescence lifetime estimations in real-time. The sensor is believed to be the first reported implementation of embedded fluorescence lifetime estimation. The system is demonstrated in a practical laboratory environment with measurements of a variety of fluorescent dyes with different single exponential lifetimes, successfully showing the sensor's ability to overcome the classic pile-up limitation of time-correlated single photon counting (TCSPC) by over an order of magnitude.

References

23