Journal of Biomedical Materials Research · 1979 · 131 citations · 6 references
Natural SciencesMaterials EngineeringMaterials ScienceMechanical Behavior Of MaterialsPolymer MaterialsTribologyCreep BehaviorCreep DeformationPolymer CharacterizationPolymer PhysicsStress-strain BehaviorWear-resistant MaterialsContact Stress EffectsCreep Of Polymer MaterialsIndentation TestingPolyethylenePressure DependenceStatic Load TestingUltra-high-molecular-weight PolyethyleneWear Rates
Wear rates of orthopaedic polyethylene in 37 degrees C water are not very contact stress dependent below 1000 psi (6.9 MPa) but above that level they accelerate substantially. The pressure dependence overall follows an exponential function. Creep in the contact pressure range of 3--17 MPa and above is a much larger factor than wear in indentation effects. For accurate measurement of wear rates by depth measurements it is necessary to permit creep to proceed under static load until it ceases before beginning the wear test.
6