Couple corrosion among alloys for skeletal prostheses

W. Rostoker, C. W. Pretzel, J O Galante

Journal of Biomedical Materials Research · 1974 · 70 citations · 18 references

Concepts

Abstract

Abstract Using a metallographic technique to identify and count corrosion pits, the tendency for passive film breakdown has been studied under conditions that simulate a crevice and a combination of two different materials. Potential corrosion couples were formed among: 316L stainless steel, cast CoCrMo alloy, wrought CoCrWNi alloy, Ti–6% Al–4% V alloy, a CoNiCrMo (multiphase) alloy and graphite. Only 316L stainless steel seems significantly prone to pitting corrosion in a crevice condition. Dissimilar metals in the crevice configuration did not seem to accelerate the corrosion of stainless steel.

References

18