Review of Scientific Instruments · 2013 · 23 citations · 12 references
EngineeringMechatronics EngineeringBiomedical OpticsMetrologyOpticsDisplacement MeasurementHolographyInterferometryDynamic HolographyHeterodyne DetectionHeterodyne InterferometryInterferometric Measurement
We report on two multipoint vibrometers with user-adjustable position of the measurement spots. Both systems are using holograms for beam deflection. The measurement is based on heterodyne interferometry with a frequency difference of 5 MHz between reference and object beam. One of the systems uses programmable positioning of the spots in the object volume but is limited concerning the light efficiency. The other system is based on static holograms in combination with mechanical adjustment of the measurement spots and does not have such a general efficiency restriction. Design considerations are given and we show measurement results for both systems. In addition, we analyze the sensitivity of the systems which is a major limitation compared to single point scanning systems.
12
Dynamic holography using pixelated light modulators
Susanne Zwick, Tobias Haist, Michael Warber et al. · Applied Optics · 2010 · 58 citations
Multi-beam laser Doppler vibrometer for landmine detection
Optical Engineering · 2006 · 43 citations
Multipoint laser vibrometer for modal analysis
William N. MacPherson, Mark Reeves, David P. Towers et al. · Applied Optics · 2007 · 34 citations