Journal of Hypertension · 2009 · 52 citations · 25 references
MedicineCardiologyBioinstrumentationPhysical ExaminationTime Series AnalysisBlood Pressure ControlBlood Pressure ManagementCardiovascular MonitoringHypertension ControlHypertension ManagementPressure MeasurementAmbulatory Blood Pressure ControlAmbulatory Physiological MonitoringBlood PressureBlood Pressure MeasurementBlood Pressure MonitoringBlood Pressure MonitorsContinuous Blood Pressure MonitoringHome Blood Pressure Monitoring
OBJECTIVES: To clarify whether a shorter interval between three successive home blood pressure (HBP) readings (10 s vs. 1 min) taken twice a day gives a better prediction of the average 24-h BP and better patient compliance. DESIGN: We enrolled 56 patients from a hypertension clinic (mean age: 60 +/- 14 years; 54% female patients). The study consisted of three clinic visits, with two 4-week periods of self-monitoring of HBP between them, and a 24-h ambulatory BP monitoring at the second visit. Using a crossover design, with order randomized, the oscillometric HBP device (HEM-5001) could be programmed to take three consecutive readings at either 10-s or 1-min intervals, each of which was done for 4 weeks. Patients were asked to measure three HBP readings in the morning and evening. All the readings were stored in the memory of the monitors. RESULTS: The analyses were performed using the second-third HBP readings. The average systolic BP/diastolic BP for the 10-s and 1-min intervals at home were 136.1 +/- 15.8/77.5 +/- 9.5 and 133.2 +/- 15.5/76.9 +/- 9.3 mmHg (P = 0.001/0.19 for the differences in systolic BP and diastolic BP), respectively. The 1-min BP readings were significantly closer to the average of awake ambulatory BP (131 +/- 14/79 +/- 10 mmHg) than the 10-s interval readings. There was no significant difference in patients' compliance in taking adequate numbers of readings at the different time intervals. CONCLUSION: The 1-min interval between HBP readings gave a closer agreement with the daytime average BP than the 10-s interval.
25
Gianfranco Parati, George S. Stergiou, Roland Asmar et al. · Journal of Hypertension · 2008 · 841 citations · Full text
Medicine, Cardiology, Nephrology +17