Natural SciencesChemistryEnvironmental SciencesAnalytical ChemistryElectroanalytical MethodsElectrochemistryElectroanalytical ChemistryElectrochemical DetectionEnvironmental AnalysisEnvironmental Analytical ChemistryHeavy Metal AnalysisWorking ElectrodesAnalytical Detection LimitsAnalytical Detection Of Heavy MetalsElectrochemical MeasurementsGlassy Carbon ElectrodesHeavy Metal DetectionPlatingSpeciation In Environmental SamplesVoltammetric Detection
A hand-held, battery-powered electroanalytical instrument utilizing anodic stripping voltammetry was designed to determine simultaneously solution concentrations of Cu2+, Pb2+ and Cd2+ in environmental samples down to 10 µg l–1, with an overall analysis time of 180 s. The instrument utilizes baseline and peak correction, a glassy carbon working electrode, in situ plating and forced convection to achieve low detection limits. Analysis is undertaken by pre-treatment of the sample with 1 ml of a solution of matrix-modifying electrolyte containing a mixture of hydrochloric acid and mercury(II) salt. The instrument and its application to the determination of heavy metals in rivers, lakes and tap water are described.
11
A silent epidemic of environmental metal poisoning?
Jerome O. Nriagu · Environmental Pollution · 1988 · 588 citations