Ultrasonic large intestine thickness determination system for low anterior resection

Genta Hiramatsu, Syoji Kobashi, Yutaka Hata, Seturo Imawaki

Conference proceedings/Conference proceedings - IEEE International Conference on Systems, Man, and Cybernetics · 2008 · 11 citations · 2 references

Concepts

Abstract

We propose a thickness determination method of the intestine with the ultrasonic probe of 15 MHz. We determine the surface point by calculating correlation coefficient between surface echo and an acquisition waveform. Next, we determine the bottom point by calculating amplitude of bottom echo, correlation with surface echo and bottom echo, and interval distance between surface echo and bottom echo. Finally, we calculate the thickness between the surface and the bottom points. As the result, we have obtained the thickness within an error rate of 5.09%.