Science · 1983 · 144 citations · 7 references
Natural SciencesBiophysicsChemistryAnalytical ChemistryBiophotonicsBiosensorsFlow CytometryFluorescence SpectroscopyLimit Of DetectionOptical BiosensingOptical BiosensorsOptical SpectroscopyAnalytical Detection LimitsConcentrationFluorescence-based AssaysFluorescent BiosensingFluorescent DyesLaser-based BiosensingLaser-induced FluorescenceLaser-induced Fluorescence Sensing
A modified flow cytometer has been used to detect attogram quantities of aqueous rhodamine 6G by laser-induced fluorescence analysis. A detection limit of 28 attograms (35,000 molecules) was obtained, nearly two orders of magnitude better than earlier measurements. The detection limit in concentration units was 1.4 x 10(-13) mole per liter. During the 1-second measurement period, the total volume sampled was 0.42 microliter. On average, only half a rhodamine 6G molecule was present in the 6-picoliter probed volume.
7