Stereotactic and Functional Neurosurgery · 2006 · 11 citations · 8 references
MedicineNuclear MedicineRadiologyDiagnostic ImagingDiagnostic RadiologyTreatment PlanningBrain ImagingComputed TomographyPositron Emission TomographyStereotactic NeurosurgeryStereotactic RadiosurgeryStereotactic SurgeryLocalization Accuracy AnalysisMultimodal Localization AccuracyPositron Emission Tomography LocalizationStereotactic LocalizationStereotactic TargetingStereotactic Treatment PlanningTarget PointsTreatment Accuracy
Computed tomography (CT), magnetic resonance imaging (MRI), and positron emission tomography (PET) provide complementary information for treatment planning in stereotactic radiosurgery. We evaluated the localization accuracy of MRI and PET compared with CT. Two kinds of phantoms applicable to the Leksell G stereotactic skull frame (Elekta, Tokyo) were developed. Deviations of measured coordinates at target points (x = 50, 100, 150; y = 50, 100, 150) were determined on different axial planes (z = 30-140 for MRI and CT study and Z = 50-120 for PET and CT study). For MRI, the deviations were no more than 0.8 mm in each direction. For PET, the deviations were no more than 2.7 mm. For both imaging modalities studied, accuracy was at or below the imaging resolution (pixel size) and should be considered useful for clinical stereotactic planning purposes.
8
Correlation of glucose consumption and tumor cell density in astrocytomas
Karl Herholz, U. Pietrzyk, Jürgen Voges et al. · Journal of neurosurgery · 1993 · 147 citations
Yukihiko Sonoda, Toshihiro Kumabe, Toshiei TAKAHASHI et al. · Neurologia medico-chirurgica · 1998 · 81 citations · Full text