Applied Spectroscopy · 1977 · 52 citations · 5 references
Natural SciencesAstronomyInstrumentationOpticsSpectroscopyInfrared InstrumentationInfrared SpectroscopyOptical DetectorsOptical InstrumentationOptical InstrumentsOptical SpectroscopyComparative Performance EvaluationFourier Transform Infrared SpectroscopyInfrared SensingInterferometric SpectroscopyMonochromatorsRelative Performance AnalysisSignal Noise ReductionSpectrometer Design
Fourier transform (FT) and grating infrared spectrophotometers are compared from a theoretical viewpoint in terms of the multiplex advantage of FT spectrometers, the relative optical throughputs, the comparative performance of the detectors used with each type of system, and other more minor parameters. The signal/noise ratios of spectra measured using commercial FT and grating spectrophotometers are compared. The relative performance varies with wavenumber in the manner forecast theoretically, although the experimental performance of the FT spectrometer is slightly worse than we predicted. Optimum applications for each type of instrument are discussed in terms of these results.
5
Fourier and Hadamard transform spectrometers: a limited comparison
M. H. Tai, Martin Harwit · Applied Optics · 1976 · 26 citations
Natural Sciences, Physics, Optics +12