Comprehensive study of the ion pumping of the noble gases

B. Čobić, G Carter, J H Leck

British Journal of Applied Physics · 1961 · 32 citations · 8 references

Concepts

Abstract

The ion pumping of five noble gases by Bayard-Alpert ionization gauges has been studied extensively, employing an ultra-high-vacuum system which could be operated statically or dynamically. The dependence of the important pumping parameters, initial pumping speed and maximum quantity of gas pumpable, upon gas composition, electrode potentials and gauge temperature have been investigated. The relationship between instantaneous pumping speed and the quantity of gas pumped has been deduced, and observations upon the recovery of gas, following sorption, at ambient and elevated temperatures has also been made.

References

8