Journal of Applied Physics · 1956 · 44 citations · 5 references
EngineeringElectrical EngineeringInstrumentationNuclear InstrumentationPlasma PhysicsAtomic IonizationIonization ChambersIonization ProcessesIonization TechniquesIonized Gas RegionsCleanupGas IonizationMaximum SpeedMetal Thin FilmsPumping
Observations have been made on the electrical cleanup of different gases in a Bayard-Alpert ionization gauge. It has been found that the ion current striking the wall of such a gauge is 5 to 10 times that reaching the ion collector. It is therefore possible to account for a maximum pumping speed of 0.1 to 0.2 l/sec due to ion pumping at the gauge walls. The maximum pumping speed observed was 0.1 l/sec for nitrogen. It was found that the maximum pumping speed for nitrogen remained the same when the inner wall of the ion gauge was covered with a thin metal deposit or Aquadag. An expression has been derived on the basis of ion pumping which describes the observed cleanup.
5
D. G. Bills, F. G. Allen · Review of Scientific Instruments · 1955 · 38 citations
The electronic clean-up of gases in sealed-off vacuum systems
R N Bloomer, M.E. Haine · Vacuum · 1953 · 31 citations