International Journal of Environmental & Analytical Chemistry · 2005 · 19 citations · 3 references
Natural SciencesChemistryToxicologyBiosensorsSensorsSpectroscopyAbsorption SpectroscopyGas SensorsLaser SpectroscopyMolecular SpectroscopyOptical SpectroscopyVolatile Organic CompoundsBreath Biomarker DetectionEthylene DetectionExhaled Breath AssessmentLipid Peroxidation MarkersPhotoacoustic Gas SensorsPhotoacoustic SpectroscopySpectroscopic DetectionTunable Diode Laser Absorption Spectroscopy
Ethylene and ethane as biomarkers for lipid peroxidation in humans can be detected in the exhaled breath by means of very sensitive gas sensors based on high-resolution molecular spectroscopy. To this purpose, a laser photoacoustic spectroscopy sensor (PAS) for online ethylene monitoring and a tunable diode laser absorption spectroscopy sensor (TDLAS) for ethane detection were developed at ENEA Frascati Molecular Spectroscopy Laboratory. The sensors have been calibrated by certified mixtures. The main advantages of the breath analysis by spectroscopic technique are the high sensitivity and the absolutely non-invasive character. During an experimental study in cooperation with Umberto I Hospital (Radiology Institute) of Rome, we were able to detect very low concentrations (under 1 ppb) of trace ethylene content in the air exhaled by patients, following X-ray therapy; results are reported in this article.
3
Estimation of peroxidative damage. A critical review.
H. Esterbauer · PubMed · 1996 · 209 citations
Ethane: a marker of lipid peroxidation during cardiopulmonary bypass in humans
Kenneth A. Andreoni, M Kazui, D. Ewen Cameron et al. · Free Radical Biology and Medicine · 1999 · 44 citations · Full text
Medicine, Physiology, Surgery +11