Medical Physics · 2004 · 29 citations · 13 references
Natural SciencesElectrical EngineeringPhysicsElectronic MaterialsPlasma PhysicsPower ElectronicsX-ray DetectionX-ray ProductionBremsstrahlungBremsstrahlung EmissionCondensersElectron BremsstrahlungX-ray DiagnosticsX-ray Line Emission
The high-voltage condensers in a polarity-inversion two-stage Marx surge generator are charged from -50 to -70 kV by a power supply, and the electric charges in the condensers are discharged to an x-ray tube after closing gap switches in the surge generator with a trigger device. The x-ray tube is a demountable diode, and the turbo molecular pump evacuates air from the tube with a pressure of approximately 1 mPa. Clean molybdenum Kalpha lines are produced using a 20 microm-thick zirconium filter, since the tube utilizes a disk cathode and a rod target, and bremsstrahlung rays are not emitted in the opposite direction to that of electron acceleration. At a charging voltage of -70 kV, the instantaneous tube voltage and current were 120 kV and 1.0 kA, respectively. The x-ray pulse widths were approximately 70 ns, and the generator produced instantaneous number of Kalpha photons was approximately 3 x 10(7) photons/cm2 per pulse at 0.5 m from the source of 3.0 mm in diameter.
13
Demonstration of a Discharge Pumped Table-Top Soft-X-Ray Laser
J. J. Rocca, Vyacheslav N. Shlyaptsev, F. G. Tomasel et al. · Physical Review Letters · 1994 · 550 citations
Natural Sciences, Physics, Optics +13
R. Germer · Journal of Physics E Scientific Instruments · 1979 · 86 citations