Analysis of the Influence of the Infusion Site on Dialyser Clearances Measured in an in vitro System Mimicking Haemodialysis and Haemodiafiltration

12

Citations

7

References

1999

Year

Abstract

The present in vitro system provided accurate and reproducible results on dialyser clearances. Our experiments confirmed previous studies on the influence of QB and QD on dialyser efficacy. Further, they show that the proportional increase in postdilutional on-line diafiltration is lesser than that previously reported. More importantly, they also show that pre-dilution infusion in high efficiency systems results in a drop in dialyser clearance compared to dialysis alone, again proportional to the infusion rate. Thus, increasing the convective flow may increase dialysis efficacy even more than increasing QD alone. However, the choice of infusion site is crucial to obtaining this benefit in small molecule clearances.