Journal of High Resolution Chromatography · 1999 · 24 citations · 9 references
Natural SciencesChemistryAnalytical ChemistryMass Spectrometric AnalysisSeparation ProcessesAnalytical ChromatographyChromatographyExtraction TechniquesGas ChromatographySolvent ExtractionVolatile Organic CompoundsDicarboxylic AcidsExtraction Of Organic CompoundsExtraction SolventsFatty AcidsSplitless InjectionVolatile Fatty AcidsVolatility
The combination of on-line methylation using trimethylsulfonium hydroxide with large volume injection of 100 μL was evaluated for the analysis of organic acids and phenols in water. Solvent split injection was applied with complete evaporation of the solvent before analytes were transferred onto the GC column. Despite complete solvent removal, losses were very low compared to conventional splitless injection even for volatile acidic compounds such as propionic acid and phenol. This is explained by intermediate formation of low volatility trimethylsulfonium salts of the analytes which were held in the injector for long evaporation times of up to 10 min, if the evaporation temperature was as low as 10°C. Using a simple liquid/liquid extraction procedure, volatile fatty acids, dicarboxylic acids, benzoic acids and phenols could be detected in 5 mL of water at concentrations of 0.04–0.1 μmol/L with GC/MS in full scan mode. Lactic, pyruvic, and also malonic acids could only be detected at higher levels because of their limited extractability from water as well as their poorer methylation yields. The method provides an easy way to sensitively detect acidic compounds of medium to high volatility in water. It was applied for screening of organic acids and phenols in batch cultures of anaerobic bacteria of which one example is shown.
9
Jacek Staniewski, Hans‐Gerd Janssen, C.A.M.G. Cramers et al. · Journal of Microcolumn Separations · 1992 · 46 citations · Full text