Review of Scientific Instruments · 1978 · 49 citations · 3 references
Natural SciencesNuclear EngineeringPhysicsInstrumentationOpticsPlasma PhysicsFusion PlasmasHigh-temperature PlasmasInterferometryLaser-plasma InteractionsPlasma ConfinementPlasma DiagnosticsDensity MeasurementsInterferometer DesignInterferometric InstrumentationInterferometric MeasurementLaser InterferometryOptical InterferometersTokamak PhysicsTokamak Plasmas
A two-wavelength plasma density interferometer utilizing a CO(2) laser and an HeNe laser is described. The interferometer is being designed for use on Doublet III, a large noncircular cross-section tokamak. The use of the two wavelengths allows the distinction between fringe shifts due to plasma density and fringe shifts due to mirror vibrations. Plasma density fringe shifts of 1/10 of a fringe shift can be measured in the presence of mirror vibrations which cause several fringe shifts. A simple digital phase comparing electronic system is used.
3
L. A. Art︠s︡imovich · Nuclear Fusion · 1972 · 321 citations