Journal of Analytical Atomic Spectrometry · 1994 · 23 citations · 15 references
Natural SciencesChemistryPhysicsAnalytical ChemistryElemental AnalysisSpectroscopyAtomic Emission SpectroscopyAtomic SpectroscopyEmission SpectroscopyAtomic Emission SpectrometryEmission Intensity AnalysisEmission Spectral AnalysisInductively Coupled Plasma AnalysisInductively Coupled Plasma Atomic Emission SpectroscopyInductively Coupled Plasma SpectrometryPlasma ChemistrySample Matrix Effects
Changes in analyte emission intensity in the presence of low concentrations of potassium chloride and phosphoric acid were studied in inductively coupled plasma atomic emission spectrometry. It has been shown that both matrices can cause significant depressant effects on the relative emission intensities of Sr, Mg, Mn, Fe, Cu and Zn under normal analytical operating conditions and that the lowering of intensity was more pronounced for KCl than for H3PO4 at the same analyte to matrix concentration ratio. The matrix effects can be corrected by increasing the power supplied to the plasma, which was found to correlate with the excitation energy of the emission line.
15