Analytical Chemistry · 2013 · 166 citations · 25 references
EngineeringBioengineeringBiomedical EngineeringBiomemsBiosensorsMicrofabricationBiomedical InstrumentationMicrofluidic BiosensorsMicrofluidic DevicesMicrofluidic SystemsMicrofluidic TransportMicrofluidicsFluid Flow PatternsMicrofluidic Diagnostic DevicesMicrofluidic SensingPaper-based Analytical DevicesPaper-based BiosensorsPaper-based Microfluidic BiosensorsPaper-based MicrofluidicsPorous Flow
We demonstrate a novel method for controlling fluid flow in paper-based devices. The method delays fluid progress through a porous channel by diverting fluid into an absorbent pad-based shunt placed into contact with the channel. Parameters to control the delay include the length and the thickness of the shunt. Using this method, reproducible delays ranging from 3 to 20 min were achieved. A simple electrical circuit model was presented and used to predict the delays in a system. Results from the model showed good agreement with experimental observations. Finally, the shunts were used for the sequential delivery of fluids to a detection zone in a point-of-care compatible folding card device using biochemical reagents for the amplified detection of the malaria protein PfHRP2.