Medical Physics · 1976 · 201 citations · 0 references
Veterinary MedicineRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyComputed TomographyComputed Tomography PhysicsRadiographic ImagingX-ray Data AnalysisX-ray DetectionX-ray DetectorsComputed Tomography ReconstructionSpatial Resolution AnalysisTomographic Noise PropagationTomographic RadiographyX-ray Computed TomographyX-ray Imaging PerformanceX-ray Imaging PhysicsX-ray Imaging Systems
Statistical fluctuations, or noise, in reconstructed tomograms are analyzed as a function of x-ray energy. The result is essentially constant for energies above 50 keV, assuming constant dose and spatial resolution, and is in close agreement with the actual performance of the EMI-Scanner (water-bag model). The noise is inversely proportional to the square of cell dimensions. For scanners without a water bag, the theoretical noise is approximately 30% less.