Journal of Instrumentation · 2012 · 10 citations · 17 references
EngineeringElectrical EngineeringBiomedical ImagingDiagnostic ImagingMolecular ImagingDetector SystemsDetector TechnologyGamma-ray DetectionMedical Imaging InstrumentationScintillation DetectorsScintillator MaterialsDetector ArraysDetector CharacterizationDetector DesignDynamic RangeExperimental ValidationMedia Saturation EffectsScintillator CrystalsTransport Simulation
This paper shows a feasibility study on the use of discrete SiPMs to read out monolithic scintillator-based detectors for use in molecular imaging applications. Monte Carlo simulations are carried out in order to evaluate the basic performance of 8 × 8 arrays of discrete SiPMs with different PDE values and compare it to conventional 64 channel multi-anode PMT (MA-PMT) readout. A detailed optical transport model was incorporated into the Monte Carlo simulation and a detector module based on a monolithic scintillator crystal of 50 × 50 × 4 mm3 coupled to a MA-PMT was built for experimental validation. The effect of the SiPM dynamic range was also investigated by including a model of the saturation effects into the optical transport simulation. The results show that a detector module based on an array of 8 × 8 discrete SiPM devices (3 × 3 mm2 and PDE > 32%) is feasible as a replacement of a 64 channel MA-PMT in order to read out large monolithic crystals for MR-compatible gamma cameras.
17
GATE: a simulation toolkit for PET and SPECT
S Jan, G. Santin, D. Strul et al. · Physics in Medicine and Biology · 2004 · 2.1K citations · Full text
A promising new scintillator: cerium-doped lutetium oxyorthosilicate
Charles L. Melcher, J.S. Schweitzer · Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment · 1992 · 309 citations
Natural Sciences, Materials Engineering, Materials Science +11