Analytical Chemistry · 2000 · 67 citations · 12 references
PharmacyChemical EngineeringElectroanalytical MethodsElectrochemical EngineeringElectrochemistryChemical SensorsElectrochemical DetectionElectrochemical SensorsFlow SensorsIon-selective ElectrodesElectrochemical BehaviorElectrochemical MeasurementsIon-selective MembranesPolymeric Membrane ElectrodesPotentiometric DetectionPotentiometric MeasurementPotentiometric SensingPotentiometric Sensors
A new kind of potentiometric chip sensor for ion-selective electrodes (ISE) based on a solvent polymeric membrane is described. The chip sensor is designed to trap the organic cocktail inside the chip and to permit sample solution to flow past the membrane. The design allows the sensor to overcome technical problems of ruggedness and would therefore be ideal for industrial processes. The sensor performance for a Ba2+-ISE membrane based on a Vogtle ionophore showed electrochemical behavior similar to that observed in conventional electrodes and microelectrode arrangements.
12
Ion-selective membrane electrodes for clinical use.
U. Oesch, Daniel Ammann, W. Simon · Clinical Chemistry · 1986 · 328 citations · Full text
Shah Convolution Fourier Transform Detection
H. John Crabtree, Martin U. Kopp, A. Manz · Analytical Chemistry · 1999 · 62 citations